Clinical trial of a new bleeding-time device
American Journal of Clinical Pathology
|October 1, 1978
Summary
A new bleeding time device, Simplate II, demonstrated comparable results to the Mielke Template in healthy volunteers and patients. Both methods showed high reproducibility, with Simplate II being simpler and faster to use.
Area of Science:
- Medical Devices
- Hematology
- Clinical Trials
Background:
- Bleeding time is a critical diagnostic parameter.
- Existing devices like the Mielke Template have limitations.
- A simpler, rapid bleeding time device is needed.
Purpose of the Study:
- To compare the performance of the Simplate II device against the Mielke Template.
- To evaluate the reproducibility and ease of use of the Simplate II.
- To assess bleeding times in healthy volunteers and patients.
Main Methods:
- A clinical trial involving 24 healthy volunteers and 28 patients.
- Simultaneous bleeding time measurements using Simplate II and Mielke Template on the same arm.
- Measurements taken before and after aspirin ingestion in healthy volunteers.
- Assessment of scarring and reproducibility.
Main Results:
- No significant differences in bleeding times between Simplate II and Mielke Template in normal subjects (pre- and post-aspirin) or patients.
- Mean pre-aspirin bleeding times: Simplate II (4.75 ± 1.42 min) vs. Mielke Template (3.65 ± 1.22 min).
- Mean post-aspirin bleeding times: Simplate II (7.86 ± 2.76 min) vs. Mielke Template (7.84 ± 2.94 min).
- Similar scarring extent and high reproducibility for both devices.
- Simplate II was simpler and faster to use.
Conclusions:
- The Simplate II device is a reliable alternative to the Mielke Template for measuring bleeding time.
- The new device offers advantages in terms of simplicity and speed.
- Clinical trials confirm the efficacy and reproducibility of Simplate II.