Intermittent mitral regurgitation: an indicator of severe Beall valve dysfunction.

R J Coralli, E R Dorney, P F Walter

    Clinical Cardiology
    |July 1, 1987
    PubMed
    Summary

    Teflon discs in Beall valves experience significant wear. Intermittent mitral regurgitation in patients with these valves indicates severe dysfunction, requiring prompt cardiac catheterization and valve replacement.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Patient discomfort following pectoral defibrillator implantation using conscious sedation.

    Pacing and clinical electrophysiology : PACE·1999
    Same author

    Comparison of three different automatic defibrillator implantation approaches: pectoral implantation using conscious sedation reduces procedure times and cost.

    Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing·1998
    Same author

    Right ventricular outflow tract tachycardia as a result of blunt chest trauma.

    Pacing and clinical electrophysiology : PACE·1998
    Same author

    Frequency analysis of human atrial fibrillation using the surface electrocardiogram and its response to ibutilide.

    The American journal of cardiology·1998
    Same author

    Anatomy of atrioventricular nodal reentry investigated by intracardiac echocardiography.

    The American journal of cardiology·1997
    Same author

    Incidence and implications of abrasion of implantable cardioverter-defibrillator leads.

    The American journal of cardiology·1997

    Area of Science:

    • Cardiovascular Surgery
    • Biomaterials Science

    Background:

    • Beall valves, utilizing Teflon discs, are prone to wear and tear.
    • Monitoring patients with Beall valves for dysfunction is clinically essential.

    Purpose of the Study:

    • To highlight the clinical significance of intermittent mitral regurgitation in patients with Beall valves.
    • To correlate this auscultatory finding with severe Teflon disc wear and abnormal valve function.

    Main Methods:

    • Case series describing 4 patients with Beall valves.
    • Clinical auscultation for mitral regurgitation.
    • Cardiac catheterization to assess valve disc function.

    Main Results:

    • All 4 patients presented with intermittent mitral regurgitation.

    Related Experiment Videos

  • Severe wear of the Teflon disc was observed in all cases.
  • Abnormal, intermittent disc tilting was confirmed via cardiac catheterization.
  • Conclusions:

    • Intermittent mitral regurgitation is a critical sign of severe Beall valve dysfunction.
    • Urgent cardiac catheterization and valve replacement are indicated upon detecting this finding.