Effects of a Fluoroscopy Agent on Radio Opacity and Steering Performance of Pressure-Driven Steerable Micro Guidewire
Abstract:
This paper presents experimental results of effects of a fluoroscopy agent on the radio opacity and steering performance of the steerable micro guidewire. The guidewire is driven by internal pressure, and made of the silicone polymer mixed with the barium sulfate, BaSO4, masterbatch. Steerable distal tips with different BaSO4 densities up to 30 % are fabricated. The radio opacity is measured by comparing CT (computed tomography) numbers of the steerable distal tips. The steering performance is measured by the bending angle at particular internal pressures while being bent up to 180 0. Experiment results show that the radio opacity improves while the bending stiffness decreases as the concentration of the barium sulfate increases.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies III: Computed Tomography