A Hardware-based Lightweight ANN for Real-time Wearable Blood Pressure Estimation
Summary
This study introduces a lightweight artificial neural network (ANN) for wearable blood pressure (BP) estimation. The developed ultra-low-power application-specific integrated circuit (ASIC) demonstrates efficient and accurate BP monitoring.
Area of Science:
- Biomedical Engineering
- Cardiovascular Health Technology
- Artificial Intelligence in Healthcare
Background:
- Blood pressure (BP) monitoring is crucial for cardiovascular health and hypertension prevention.
- There is a growing demand for low-power wearable devices for continuous BP estimation.
- Existing methods often face challenges in power consumption and miniaturization for wearable applications.
Purpose of the Study:
- To propose and implement a lightweight artificial neural network (ANN) for accurate and low-power wearable blood pressure (BP) estimation.
- To integrate the ANN model onto an ultra-low-power application-specific integrated circuit (ASIC).
- To evaluate the performance and power efficiency of the developed ASIC for BP monitoring.
Main Methods:
- Development of a lightweight artificial neural network (ANN) model tailored for BP estimation.
- Implementation of the ANN model on an ultra-low-power application-specific integrated circuit (ASIC) using a 0.18 µm CMOS process.
- Validation of the ASIC's performance using standard metrics like Mean Absolute Error (MAE), Standard Deviation (SD), and R-squared (R²) for systolic and diastolic BP estimation.
Main Results:
- The ASIC achieved a low power consumption of 19.72 µW and occupied an area of 730 µm x 730 µm.
- On the test set, the estimated systolic and diastolic BP showed MAE ± SD of 2.47 ± 3.48 mmHg and 1.45 ± 1.88 mmHg, respectively.
- With 8-bit quantization, the MAE ± SD for systolic and diastolic BP were 12.41 ± 5.32 mmHg and 6.29 ± 3.03 mmHg, with an overall R² of 0.9702.
Conclusions:
- The proposed lightweight ANN implemented on an ultra-low-power ASIC offers a viable solution for efficient wearable blood pressure estimation.
- The developed ASIC demonstrates high accuracy and low power consumption, suitable for continuous monitoring applications.
- This technology contributes to advancements in non-invasive hypertension management and personalized cardiovascular health tracking.
Related Concept Videos
Equipments Used To Measure Blood Pressure
1.2K
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
1.2K
Measurement of Blood Pressure
1.3K
Assessing blood pressure is a standard procedure executed in virtually all medical environments. The method utilized today was established over a hundred years ago by an innovative Russian doctor, Dr. Nikolai Korotkoff. The soft ticking noise, known as Korotkoff sounds, heard while taking blood pressure readings results from turbulent blood flow within the vessels. The apparatus required for this procedure includes a sphygmomanometer, a blood pressure cuff attached to a gauge, and a...
1.3K
Pre-Procedural Guidelines for Assessing Blood Pressure
621
Accurate blood pressure assessment is crucial for diagnosing and managing various health conditions. To ensure the reliability of these measurements, healthcare professionals must adhere to standardized pre-procedural guidelines. These guidelines enhance patient safety and improve the overall quality of healthcare. The following steps are essential for obtaining accurate and consistent blood pressure readings, from using the appropriate tools to ensuring effective communication with the...
621
Neural Regulation of Blood Pressure
3.3K
The neural regulation of blood pressure involves intricate interactions between the autonomic nervous system (ANS) and cardiovascular system, ensuring adequate perfusion of tissues. This regulation primarily occurs through baroreceptor and chemoreceptor reflexes, involving both short-term and long-term mechanisms.
Baroreceptor Reflex
Baroreceptors, located in the carotid sinuses and aortic arch, detect changes in blood pressure. When blood pressure rises, these stretch-sensitive receptors...
Baroreceptor Reflex
Baroreceptors, located in the carotid sinuses and aortic arch, detect changes in blood pressure. When blood pressure rises, these stretch-sensitive receptors...
3.3K
Assessing Blood pressure using a doppler ultrasound
1.6K
To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
1.6K
Sites for measruring blood pressure
2.3K
Blood pressure measurement is a fundamental clinical procedure, providing crucial data for assessing cardiovascular health. Among the various sites for this measurement, the brachial and popliteal arteries are predominantly utilized due to their accessibility and the reliability of their readings. This lesson delves into the anatomical significance, methodology, and considerations of measuring blood pressure at these locations.
The Brachial Artery: Primary Site for Blood Pressure Measurement
The Brachial Artery: Primary Site for Blood Pressure Measurement
2.3K