Quantitative detection of seizures with minimal-density EEG montage using phase synchrony and cross-channel coherence
Insights
Detecting seizures in critically ill children is challenging. This study shows a new method using 4-8 electroencephalogram (EEG) channels can detect seizures with 77% accuracy, enabling bedside diagnosis.
Area of Science:
- Biomedical Engineering
- Clinical Neurology
- Computational Neuroscience
Background:
- Sub-clinical seizures are common in pediatric critical care, complicating diagnosis and potentially worsening outcomes.
- Current gold-standard seizure detection via electroencephalograms (EEG) requires specialized personnel for electrode placement and interpretation.
- Delays in seizure detection and treatment negatively impact neurological outcomes in critically ill children.
Purpose of the Study:
- To develop and validate an automated seizure detection algorithm using limited EEG channels.
- To assess the feasibility of bedside seizure detection independent of expert input.
- To evaluate the accuracy of phase synchrony and cross-channel coherence for seizure identification.
Main Methods:
- Extracted phase synchrony and cross-channel coherence from 4 and 8 pre-selected scalp EEG signals.
- Employed binary classification to differentiate seizure from non-seizure signal segments.
- Compared algorithm predictions against gold-standard seizure onset markings from multi-channel EEGs.
Main Results:
- The algorithm achieved an average accuracy of 77.2% using 4 EEG channels.
- An average accuracy of 76.5% was obtained using 8 EEG channels.
- Demonstrated the feasibility of seizure detection with a simplified 4 or 8-channel EEG montage.
Conclusions:
- Automated seizure detection is feasible using a limited number of EEG channels (4 or 8).
- This approach allows for readily applicable bedside seizure detection, reducing reliance on expert interpretation.
- The findings support the potential for earlier and more accessible seizure detection in pediatric critical care settings.
Abstract:
Seizures frequently occur in paediatric emergency and critical care, with up to 74% being sub-clinical seizures making detection difficult. Delays in seizure detection and treatment worsen the neurological outcome of critically-ill patients. Gold-standard seizure detections using multi-channels electroencephalograms (EEG) require trained clinical physiologists to apply scalp electrodes and highly specialised neurologists to interpret and identify seizures. In this study, we extracted phase synchrony and cross-channel coherence amplitude across 4 and 8 pre-selected scalp EEG signals. Binary classification is used to determine whether the signal segment is seizure or non-seizure, and the predictions were compared against the gold-standard seizure onset markings. The application of the algorithm on a cohort of forty routinely collected EEGs from paediatric patients showed an average accuracy of 77.2 % and 76.5% using 4 and 8 channels, respectively. Clinical Relevance- This work demonstrates the feasibility of seizure detection with pre-defined 4 and 8 EEG electrodes with an average accuracy of 77%. This means for the first time seizure detection is possible using an EEG montage that can be applied readily at the bedside independent of expert input.