Related Experiment Video
Updated: Aug 29, 2025

Effective Analysis of Human Exposure Conditions with Body-worn Dosimeters in the 2.4 GHz Band
Published on: May 2, 2018
Design of Flexible Meander Line Antenna for Healthcare in Wireless Body Area Network Systems
Abstract:
A flexible meander line monopole antenna (MMA) is presented in this paper. The antenna can be worn for on-and off-body applications. The overall dimension of the MMA is 37 mm x 50 mm x 2.37 mms, The MMA was manufactured and measured, and the results matched with simulation results. The MMA design shows a bandwidth of up to 1282.4 (450.5) MHz and provides gains of 3.03 (4.85) dBi in two operating bands, respectively, showing omnidirectional radiation patterns in free space. While worn on the chest or arm, bandwidths as high as 688.9 (500.9) MHz and 1261.7 (524.2) MHz, and the gains of 3.80 (4.67) dBi and 3.00 (4.55) dBi were observed. The experimental measurements of the read range confirmed. Clinical Relevance- Wireless Medical Body Area Network (WMBAN) technology allows for continuous monitoring and analysis of patient health data to improve the quality of healthcare services.
More Related Videos
08:25Construction of a Wireless-Enabled Endoscopically Implantable Sensor for pH Monitoring with Zero-Bias Schottky Diode-based Receiver
Published on: August 27, 2021
08:50Fabrication and Characterization of a Conformal Skin-like Electronic System for Quantitative, Cutaneous Wound Management
Published on: September 2, 2015