NANOCIS®: Possible errors in the summary of product characteristics quality control method and a suggested
J P Paludan1, R Kraack2, L Falborg3
1Department of Clinical Physiology, Regional Hospital Viborg, Denmark.
Background:
Determination of the radiochemical purity of [99mTc]NANOCIS® was performed using the procedure described in the Summary of Product Characteristics (SPC). In contrast to the clinical findings of thyroid gland accumulation indicating free [99mTc] pertechnetate, the QC results showed no free [99mTc]pertechnetate. This discrepancy prompted us to further investigate the described QC procedure. The aim of our study is to develop a correct QC procedure for [99mTc]NANOCIS®.
Method And Materials:
After 99mTc-labelling performed in accordance with the SPC, QC was performed on two stationary phases (Whatman No. 1 and ITLC-SA) with both wet and dry application spots.
Results:
All QC samples prepared using the method described in the SPC (Whatman No. 1 with dry application spot) indicated an acceptable labelling with a radiochemical purity over 99 %. The QC methods performed using non-SPC described methods (Whatman No. 1 with wet application spot, ITLC-SA with wet and dry application spot), show more impurities, resulting in radiochemical purity ranging between 68 % and 99 %. All results from the three QC procedures not outlined in the SPC resulted in comparable results. When comparing the QC results with imaging of the thyroid gland, if the correct TLC method is used, a clear connection was observed between low radiochemical purity, as a result of free [99mTc]pertechnetate present in the prepared radiopharmaceutical, and visualisation of the thyroid gland, due to thyroid uptake of the free [99mTc]pertechnetate.
Conclusion:
Drying the application spot on Whatman No. 1 paper indicates erroneous high labelling. To obtain a correct QC result, either an analytical method using Whatman No. 1 papers without drying of the application spot or a method using instant thin layer chromatography with a silica acid coating (ITLC-SA) should be used.
More Related Videos
10:22Split Point Analysis and Uncertainty Quantification of Thermal-Optical Organic/Elemental Carbon Measurements
Published on: September 7, 2019
09:24Application and Methodology of the Non-destructive 19F Time-domain NMR Technique to Measure the Content in Fluorine-containing Drug Products
Published on: August 22, 2017
Related Concept Videos
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...
Quality Assurance
Data Validation
Key parameters for method validation include:
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...