Related Experiment Video
Updated: Jul 30, 2026

Large Scale Non-targeted Metabolomic Profiling of Serum by Ultra Performance Liquid Chromatography-Mass Spectrometry UPLC-MS
Published on: March 14, 2013
A Second-generation opioid LC-MS/MS assay improves laboratory workflow and capacity.
Hsuan-Chieh Liao1, Thomas J Laha1, Geoffrey S Baird1
1Department of Laboratory Medicine and Pathology, University of Washington, Seattle, WA.
A new "dilute-and-shoot" liquid chromatography-tandem mass spectrometry (LC-MS/MS) assay simplifies opioid drug testing. This method improves sensitivity, reduces errors, and doubles laboratory testing capacity for addiction and pain management.
Area of Science:
- Clinical Chemistry
- Analytical Chemistry
- Forensic Toxicology
Background:
- Opioid drug misuse is a significant public health crisis, driving the need for accurate and efficient diagnostic tools.
- Urine drug screening is crucial for managing patients with chronic pain or opioid use disorder.
- Liquid chromatography-tandem mass spectrometry (LC-MS/MS) is the gold standard for confirmatory drug testing due to its high sensitivity and specificity, but its complexity poses challenges.
Purpose of the Study:
- To develop and validate a simplified "dilute-and-shoot" LC-MS/MS assay for opioid testing.
- To leverage upgraded instrumentation to overcome operational complexities of LC-MS/MS.
- To improve assay performance metrics including sensitivity, dynamic range, and turnaround time.
Main Methods:
- Development of a simplified "dilute-and-shoot" LC-MS/MS method for opioid analysis.
- Optimization of chromatographic gradients to resolve isobaric interferences.
- Expansion of analytical ranges using alternative mass transitions and updated quality assurance parameters.
- Validation using 204 clinical samples, comparing results with a previous assay version.
Main Results:
- The new LC-MS/MS method demonstrated good correlation with the previous assay.
- Upgraded instrumentation resulted in enhanced sensitivity and expanded dynamic ranges.
- The simplified assay reduced carryover, eliminating the need for extra injections and chromatogram reviews.
- Turnaround time was reduced, and testing capacity was effectively doubled.
Conclusions:
- The developed "dilute-and-shoot" LC-MS/MS assay offers a streamlined and efficient approach to opioid drug testing.
- This method enhances laboratory efficiency by improving sensitivity, reducing manual review, and increasing throughput.
- The improvements serve as a valuable model for clinical laboratories transitioning to new mass spectrometry instrumentation.