Prevalence of bacterial burden on macroscopic contaminants of orthopaedic surgical instruments following sterilization
Related Experiment Video
Updated: Aug 29, 2025

Novel Diagnostics in Revision Arthroplasty: Implant Sonication and Multiplex Polymerase Chain Reaction
Published on: December 3, 2017
I J Wellington1, T J Schneider1, B C Hawthorne1
1Department of Orthopedics, University of Connecticut, Farmington, CT, USA.
Background:
Macroscopic contamination of orthopaedic instruments with particulates, including cortical bone and polymethyl methacrylate (PMMA) bone cement, that have previously undergone pre-operative sterilization is frequently encountered peri- or intraoperatively, calling into question the sterility of such instruments.
Aim:
To determine if macroscopic contaminants of orthopaedic surgical instrumentation maintain a bacterial burden following sterile processing, and to determine the most commonly contaminated instruments and the most common contaminants.
Methods:
Macroscopic contaminants in orthopaedic instrument trays were collected prospectively at a single tertiary referral centre over a 6-month period from August 2021 to May 2022. When identified, these specimens were swabbed and plated on sheep blood agar. All specimens were incubated at 37 °C for 14 days, and inspected visually for colony formation. When bacterial colony formation was identified, samples were sent for species identification.
Results:
In total, 33 contaminants were tested, and only one contaminant was found to be growing bacterial colonies (Corynebacterium sp.). The items most commonly found to have macroscopic contamination were surgical trays (N=9) and cannulated drills (N=7). The identifiable contaminants were bone (N=10), PMMA bone cement (N=4) and hair (N=4). Eleven macroscopic contaminants were not identifiable.
Conclusion:
This study found that 97% of macroscopic orthopaedic surgical instrument contaminants that underwent sterile processing did not possess a bacterial burden. Contaminants discovered during a procedure are likely to be sterile, and do not pose a substantially increased risk of infection to a patient.
Insights
Most macroscopic contaminants on orthopaedic instruments, including bone and bone cement, are sterile after processing. This suggests that intraoperative contamination does not significantly increase patient infection risk.
Area of Science:
- Orthopaedic surgery
- Infection control
- Sterilization processes
Background:
- Macroscopic contamination of orthopaedic instruments with bone and bone cement is common.
- This contamination raises concerns about instrument sterility after pre-operative sterilization.
Purpose of the Study:
- To assess bacterial burden on macroscopic contaminants after sterile processing.
- To identify commonly contaminated instruments and contaminant types.
Main Methods:
- Prospective collection of macroscopic contaminants from orthopaedic instrument trays over 6 months.
- Swabbing and culturing contaminants on sheep blood agar.
- Incubation for 14 days with visual inspection and species identification if growth occurred.
Main Results:
- 33 contaminants were tested; only one (Corynebacterium sp.) showed bacterial growth.
- Surgical trays (N=9) and cannulated drills (N=7) were most frequently contaminated.
- Bone (N=10) and polymethyl methacrylate (PMMA) bone cement (N=4) were common identifiable contaminants.
Conclusions:
- 97% of macroscopic contaminants on orthopaedic instruments lacked a bacterial burden post-sterilization.
- Intraoperative contaminants are likely sterile and do not pose a significant infection risk.
- Findings support the safety of using instruments with observed macroscopic contamination after sterile processing.
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...