Related Experiment Video
Updated: Aug 29, 2025

Three-Dimensional Preoperative Virtual Planning in Derotational Proximal Femoral Osteotomy
Published on: February 17, 2023
Accuracy of intraoperative bone registration and stereotactic boundary reconstruction during total knee arthroplasty
Stijn Herregodts1,2, Hannes Vermue1, Jan Herregodts1
1Department of Human Structure and Repair, Ghent University, Ghent, Belgium.
Background:
The intraoperative registration of the bones play a crucial role in image-based computer-assisted knee arthroplasty to achieve accurate implant placement and to create reliable stereotactic bone boundaries for robot-assisted surgical systems.
Method:
This study assessed the intraoperative registration accuracy on six intact fresh frozen cadavers.
Results:
Rotational errors around the mechanical axis were the largest, with a standard deviation of 1.2° and outliers up to 3.7°. The mean translational errors were lower than 1 mm, with outliers up to 1.5 mm. These errors were amplified to 2 mm for the registration-based reconstruction of the posterior bone surface at the resection levels.
Conclusion:
Given the cumulative behaviour of surgical errors, registration errors can affect the final implant positioning. Furthermore, inaccuracies in the reconstructed bone boundary directly affect the virtual stereotactic boundaries used in robotic-assisted surgery and can result in an incomplete resection or inadvertent soft tissue damage.