Improving reliability for measuring ST deviation in suspected acute coronary syndrome
Yu-He Zhang1, Richard Gregg1, Saeed Babaeizadeh1
1Advanced Algorithm Research Center, Philips Healthcare, Cambridge, MA, USA.
Background:
Standard 12‑lead ECG is used for diagnosis and risk stratification in suspected acute coronary syndrome (ACS) patients. Artifacts have significant impact on the measuring quality, which consequently affect the diagnostic decision. We used a signal quality indicator (SQI) to identify the ECG segments with lower artifact levels which we hypothesized would improve ST measurements.
Methods:
The Staff III 12‑lead ECG database was used with the ECG segments before balloon inflation (n = 185). SQI scores per second were calculated and a 10-s ECG segment with least noise and artifacts (Clean10) was identified for each minute of recording. The first 10 s of ECG recordings (First10) for each minute were selected as a reference. The Philips DXL™ algorithm was used to measure the ST levels at J-point, +20 ms, +40 ms, +60 ms, and + 80 ms after the J-point. Standard deviations (SDs) for the ST measurements for each of the 185 ECG records were calculated for the Clean10 and for the First10 across records. The resulting SDs for the Clean10 were compared with the SDs for the First10 using the Wilcoxon signed rank test.
Results:
The results indicated that 1) The SDs for the Clean10 are lower than that of the First10; 2) The SDs for J+20 ms and J+40 ms are lowest among the 5 different measuring points although similar improvement for the Clean10 over the First10 is observed for J+60 ms and J+80 ms as well; 3) The improvement at the J-point was not as high as other ST measurements.
Conclusions:
The SQI is demonstrated as an efficient tool to identify the ECG segments with lower artifacts that produce more consistent and reliable ST measurement. The measurements at J+20 ms demonstrated the highest consistency among the five studied measuring points.
More Related Videos
10:01Improvement of a Closed Chest Porcine Myocardial Infarction Model by Standardization of Tissue and Blood Sampling Procedures
Published on: March 12, 2018
06:16Signal Acquisition, Score Interpretation, and Economics of a Non-Invasive Point-of-Care Test for Coronary Artery Disease
Published on: August 9, 2024
Related Concept Videos
Acute Coronary Syndrome III: Diagnostic Studies
Testing a Claim about Standard Deviation
The hypothesis testing for the claim of population standard deviation (or variance) requires the data and samples to be random and unbiased. The population distribution also must be normal. There is no specific requirement on the sample size as the estimation is based on the chi-square distribution.
As a first step, the hypothesis (null and alternative) concerning the claim about...
Acute Coronary Syndrome II: Pathophysiology and Clinical Manifestations
Errors occurring during blood pressure monitoring
Several factors...
Acute Coronary Syndrome I: Introduction
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT