Objectives:
To evaluate the feasibility, accuracy, usability and acceptability of two upper arm self-sampling devices for measurement of autoantibodies and C reactive protein (CRP) levels in patients with immune-mediated rheumatic diseases (IMRDs).
Methods:
70 consecutive patients with IMRD with previously documented autoantibodies were assigned to supervised and unsupervised self-collection of capillary blood with the Tasso+ or TAP II device. Interchangeability of 17 biomarkers with standard venesection was assessed by: concordance, correlation, paired sample hypothesis testing and Bland-Altman plots. Patients completed an evaluation questionnaire, including the System Usability Scale (SUS) and Net Promoter Score (NPS).
Results:
While 80.0% and 77.0% were able to safely and successfully collect capillary blood using the Tasso+ and TAP II within the first attempt, 69 of 70 (98.6%) patients were successful in collecting capillary blood within two attempts. Concordance between venous and capillary samples was high; 94.7% and 99.5% for positive and negative samples, respectively. For connective tissue disease screen, anti-Ro52 and anti-proteinase 3 autoantibody levels, no significant differences were observed. Self-sampling was less painful than standard venesection for the majority of patients (Tasso+: 71%; TAP II: 63%). Both devices were well accepted (NPS; both: +28%), usability was perceived as excellent (SUS; Tasso+: 88.6 of 100; TAP II: 86.0 of 100) and 48.6 %/62.9% of patients would prefer to use the Tasso+/TAP II, respectively, instead of a traditional venous blood collection.
Conclusions:
Remote self-collection of capillary blood using upper arm-based devices for autoantibody and CRP analysis in patients with autoimmune rheumatic diseases is feasible, accurate and well accepted among patients.
Trial Registration Number:
WHO International Clinical Trials Registry (DRKS00024925).