A Continuous Pacing and Recording Technique for Differentiating Left Bundle Branch Pacing From Left Ventricular
Jiabo Shen1, Longfu Jiang1, Hao Wu1
1Department of Cardiology, Hwa Mei Hospital, University of Chinese Academy of Sciences, Ningbo, Zhejiang, China.
Background:
Left bundle branch pacing (LBBP) is a promising approach for achieving near-physiologic pacing. However, differentiating LBBP from left ventricular septal endocardial pacing (LVS(e)P) remains a challenge. This study aimed to establish a simple and effective method for differentiating LBBP from LVS(e)P and to evaluate their electrophysiologic characteristics.
Methods:
LBBP, using continuous uninterrupted pacing and real-time monitoring of electrocardiograms along with intracardiac electrograms, was performed in 97 consecutive patients. We evaluated the electrophysiologic characteristics observed during LBBP using 6 modalities: right ventricular septal pacing (RVSP), intraventricular septal pacing (IVSP 1 and 2), LVS(e)P, nonselective LBBP (NSLBBP), and selective LBBP (SLBBP).
Results:
Of the 97 patients, 87 (89.7%) met the criteria (abrupt change in paced QRS morphology with a transition from Qr to QR/qR in lead V1 and shortening of stimulus to V6 R-wave peak time [Stim-V6RWPT] of ≥ 10 ms with constant output while rather than after lead screwing) for nonselective left bundle branch (LBB) capture. Selective LBB capture was observed in 82 patients (84.5%). The Stim-V6RWPT of NSLBBP and SLBBP were significantly shorter than LVS(e)P (respectively, 67.1 ± 8.7 ms, 67.0 ± 9.3 ms, and 82.1 ± 10.9 ms). Stim-QRSend was the narrowest in IVSP2 (136.6 ± 15.2 ms) instead of NSLBBP (140.0 ± 17.1 ms).
Conclusions:
The uninterrupted pacing technique for differentiating LBBP from LVS(e)P in the same group of patients is feasible. Electrophysiologic evidence from our intrapatient-controlled study shows that LBBP and LVS(e)P differ in ventricular electrical synchronization.