Measurement of Urinary Ammonium Using a Commercially Available Plasma Ammonium Assay
Valentinas Gruzdys1, Kenneth Cahoon2, Lauren Pearson3
1Doctor's Data, St. Charles, Illinois.
A modified plasma ammonium assay accurately measures urinary ammonium excretion, offering a more accessible method for diagnosing kidney and acid-base disorders. This assay demonstrates stability and reliability for clinical use.
Area of Science:
- Clinical Chemistry
- Nephrology
- Biochemistry
Background:
- Urinary ammonium excretion is crucial for evaluating acid-base disorders, chronic kidney disease, and nephrolithiasis.
- Current measurement methods are limited to specialized laboratories in the United States, hindering widespread clinical application.
- This study aimed to adapt a plasma ammonium assay for quantifying urinary ammonium and assess its stability.
Purpose of the Study:
- To evaluate the performance of a modified enzymatic plasma ammonium assay for measuring urinary ammonium concentration.
- To determine the stability of urinary ammonium under various storage conditions (temperature, pH).
- To assess the suitability of the modified assay for routine clinical use.
Main Methods:
- An enzymatic plasma ammonium assay was modified for urine samples by employing a 40-fold dilution.
- Assay precision, limit of quantitation, and linearity were evaluated using an Abbott Architect ci8200 analyzer.
- Method comparison with the formalin titration method and stability studies at different temperatures and pH levels were conducted.
Main Results:
- The modified assay demonstrated good precision (2-18%) and linearity (0.7-45 mmol/L) with a limit of quantitation of 2.92 mmol/L.
- Comparison with the formalin method showed strong agreement (slope 0.98, intercept -0.37 mmol/L).
- Urinary ammonium was stable for 48 hours at room temperature, 9 days at 4°C/-20°C, and over 24 months at -80°C within a pH range of 5.6-6.3.
Conclusions:
- The modified enzymatic plasma ammonium assay reliably quantifies urinary ammonium at physiologic concentrations.
- The assay demonstrates excellent performance characteristics and stability, comparable to the established formalin titration method.
- This adapted assay is suitable for routine clinical use on automated chemistry analyzers, improving accessibility for patient diagnostics.