Background:
Right ventricular (RV) apical pacing can induce both interventricular dyssynchrony and intraventricular dyssynchrony. Mechanical dyssynchrony after long-term RV apical pacing is associated with reduced left ventricular (LV) systolic function and deterioration in functional capacity.
Aim:
We aimed to identify the short-term effects of the pacemaker RV lead position on remodeling of LV systolic functions.
Patients And Methods:
The study included 30 patients who presented with an indication of permanent pacing and who underwent permanent single- or dual-chamber pacemaker insertion: 15 patients with RV apical pacing (RV apex), and 15 patients with non-apical pacing (mid-septal). The two-dimensional (2D) speckle tracking imaging technique was used for quantification of global longitudinal function of the left ventricle and dyssynchrony evaluation before pacemaker implantation and after a 3-month follow-up.
Results:
At the 3‑month follow-up, post-pacing 2D speckle tracking echocardiography revealed impairment of global longitudinal strain in all patients and intraventricular dyssynchrony was significantly increased in the apical location compared with the non-apical location (radial dyssynchrony: 108.67 ± 11.68 ms vs. 80.53 ± 8.17 ms, p < 0.001) with a greater difference (50.53 ± 13.30 ms) in the apical location than in the non-apical location (29.87 ± 6.64 ms, p < 0.001).
Conclusion:
In the short-term follow-up, 2D speckle tracking echocardiography showed more radial dyssynchrony in the apical location than in the non-apical location of RV lead. The RV septal pacing is a better alternative in terms of less dyssynchrony compared to RV apical pacing. Older age, higher percentage of pacing, and device type are prognostic factors for development of pacemaker-induced cardiomyopathy.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
10:39Rat Model of Right-Sided Cardiac Remodeling and Arrhythmia Using Pulmonary Artery Banding
Published on: August 30, 2024
Related Concept Videos
Mitral Regurgitation I: Introduction
Mitral Stenosis I: Introduction