Related Experiment Video
Updated: Aug 28, 2025

Neonatal Subventricular Zone Electroporation
Published on: February 11, 2013
Bipolar Electroporation Across the Interventricular Septum: Electrophysiological, Imaging, and Histopathological
Martin van Zyl1, Thomas P Ladas1, Jason A Tri1
1Department of Cardiovascular Medicine, Mayo Clinic College of Medicine and Science, Rochester, Minnesota, USA.
Background:
Pulsed electric field (PEF) ablation is an emerging modality for the treatment of cardiac arrhythmias. Data regarding effects on the interventricular septum are limited, and the optimal delivery protocol and electrode configuration remain undefined.
Objectives:
This study sought to evaluate the electrophysiological, imaging, and histological characteristics of bipolar direct-current PEF delivered across the interventricular septum.
Methods:
PEF was applied between identical solid-tip ablation catheters positioned on either side of the septum in a chronic canine model. Intracardiac and surface electrophysiological data were recorded following delivery. In 4 animals, cardiac magnetic resonance (CMR) was performed early (6 ± 2 days) and late (30 ± 2 days) postablation. After 4 weeks of survival, cardiac specimens were sectioned for histopathological analysis.
Results:
In 8 canines, PEF was delivered in 27 separate septal sites (45 ± 17 J/site) with either microsecond or nanosecond PEF. Acute complications included transient complete atrioventricular block in 5 animals (63%) after delivery at the anterobasal septum, with right bundle branch block persisting in 3 (38%). Ventricular fibrillation occurred in 1 animal during microsecond but not nanosecond PEF. Postprocedural CMR showed prominent edema and significant left ventricular systolic dysfunction, which recovered with late imaging. At 4 weeks, 36 individual well-demarcated lesions were demonstrated by CMR and histopathology. Lesion depth measured by histology was 2.6 ± 2.1 mm (maximum 10.9 mm and near transmural).
Conclusions:
Bipolar PEF ablation of the interventricular septum is feasible and can produce near transmural lesions. Myocardial stunning, edema, and conduction system injury may occur transiently. Further studies are required to optimize safe delivery and efficacious lesions.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
10:12Monitoring Electroporation-Induced Changes in Action Potential Generation in Genetically Engineered Tet-On Spiking HEK cells
Published on: September 6, 2024