Influence of Abutment Design on Biomechanical Behavior to Support a Screw-Retained 3-Unit Fixed Partial Denture
Guilherme da Rocha Scalzer Lopes1, Jefferson David Melo de Matos1,2, Daher Antonio Queiroz3
1Department of Biomaterials, Dental Materials and Prosthodontics, Institute of Science and Technology, São Paulo State University (Unesp), São José dos Campos 12245-000, Brazil.
This study found no significant biomechanical difference between CMN and miniconical abutments for supporting fixed partial dentures. Both abutment types are suitable for screw-retained dental restorations under favorable conditions.
Area of Science:
- Biomaterials Science
- Dental Implantology
- Biomechanics
Background:
- Dental implants are crucial for supporting fixed partial dentures.
- Abutment selection impacts the biomechanical behavior of implant-supported prostheses.
- Morse taper implants offer a stable connection, but abutment design variations warrant investigation.
Purpose of the Study:
- To compare the biomechanical performance of Morse taper implants with CMN abutments versus miniconical abutments.
- To evaluate stress and strain distribution in screw-retained 3-unit fixed partial dentures supported by different abutments.
- To validate finite element analysis (FEA) models using in vitro strain gauge data.
Main Methods:
- In vitro testing using polyurethane blocks with Morse taper implants and either CMN or miniconical abutments.
- Strain gauge measurements were taken under a 300 N axial load.
- 3D finite element analysis was performed to simulate static structural behavior.
Main Results:
- In vitro strain gauge analysis showed no statistically significant difference in microstrain between CMN (645.3 ± 309.2 με) and MC (639.3 ± 278.8 με) groups (p = 0.879).
- FEA results also indicated no significant difference in microstrain between CMN (605.1 ± 358.6 με) and MC (598.7 ± 357.9 με) groups (p = 0.932).
- Computational models demonstrated good agreement with experimental data, validating the FEA approach.
Conclusions:
- Both CMN and miniconical abutments demonstrate comparable biomechanical behavior when supporting a 3-unit fixed partial denture on Morse taper implants.
- The choice between CMN and miniconical abutments may not significantly affect the mechanical stability of such restorations under optimal conditions.
- These findings support the clinical indication of both CMN and miniconical abutments for screw-retained fixed partial dentures in favorable scenarios.