Abstract:
Advancements in digital imaging technologies hold the potential to transform prosthetic and orthotic practices. Non-contact optical scanners can capture the shape of the residual limb quickly, accurately, and reliably. However, their suitability in clinical practice, particularly for the transradial (below-elbow) residual limb, is unknown. This project aimed to evaluate the reliability of an optical scanner-based shape capture process for transradial residual limbs related to volumetric measurements and shape assessment in a clinical setting. A dedicated setup for digitally shape capturing transradial residual limbs was developed, addressing challenges with scanning of small residual limb size and aspects such as positioning and patient movement. Two observers performed three measurements each on 15 participants with transradial-level limb absence. Overall, the developed shape capture process was found to be highly repeatable, with excellent intra- and inter-rater reliability that was comparable to the scanning of residual limb cast models. Future work in this area should compare the differences between residual limb shapes captured through digital and manual methods.