Inter-fractional entrance dose monitoring as quality assurance using Gafchromic EBT3 film
Related Experiment Video
Updated: Aug 27, 2025

Dosimetry for Cell Irradiation using Orthovoltage 40-300 kV X-Ray Facilities
Published on: February 20, 2021
Sun Young Moon1, Yunhui Jo2, Jaehyeon Seo2
1Department of Bio-convergence Engineering, Korea University, Seoul; Proton Therapy Center, National Cancer Center, Goyang, Korea.
Introduction:
This study describes a simple method of inter-fractional photon beam monitoring to measure the entrance dose of radiation treatment using Gafchromic EBT3 film.
Materials And Methods:
The film was placed at the center of a 1-cm thick phantom shaped like a block tray and fixed on the accessory tray of the gantry. The entrance dose was measured following the placement of the film in the accessory tray. The dose distribution calculated with the treatment planning system was compared with the dose distribution on the irradiated EBT3 films. The effectiveness of this methodology, as determined by gamma passing rates, was evaluated for the 22 fields of eight three-dimensional conformal radiotherapy (3D-CRT) plans and the 41 fields of nine intensity-modulated radiotherapy (RT) plans. The plans for three-dimensional conformal RT included treatments of the rectum, liver, breast, and head and neck, whereas the plans for intensity-modulated RT included treatments of the liver, brain, and lung.
Results:
The gamma passing rates for 3D-CRT ranged from 96.4% to 99.5%, with the mean gamma passing rate for 22 fields being 98.0%. The gamma passing rate for intensity-modulated RT ranged from 96.1% to 98.9%, with the mean gamma passing rate for 41 fields being 97.7%. All gamma indices were over the 95% tolerance level.
Conclusions:
The methodology described in this study, based on Gafchromic EBT3 film, can be utilized for inter-fractional entrance dose monitoring as quality assurance during RT. Clinical application of this method to patients can verify the accuracy of beam delivery in the treatment room.