Related Experiment Video
Updated: Aug 27, 2025

Phthalic Acid Ester-Binding DNA Aptamer Selection, Characterization, and Application to an Electrochemical Aptasensor
Published on: March 21, 2018
Molecularly imprinted polypyrrole based electrochemical sensor for selective determination of ethanethiol
M Asunción Alonso-Lomillo1, Olga Domínguez-Renedo1
1Analytical Chemistry Department, Faculty of Sciences, University of Burgos, Pza. Misael Bañuelos S/n, 09001, Burgos, Spain.
Abstract:
This work describes a molecularly imprinted (MIP) sensor, based on the electropolymerization of pyrrole on a glassy carbon electrode (GCE), for the determination of ethanethiol. Ethanethiol was used as a template molecule for the formation of cavities in the imprinted polymer. The effect of molar ratios template molecules/functional monomers and time needed to remove the template were optimized. The developed MIP/GCE sensor presented a linear range from 6.1 to 32.4 mg L-1 with capability detection and reproducibility values of 7.2 mg L-1 and 10.4%, respectively. The sensitivity of the developed sensor was enhanced by the incorporation of gold nanoparticles (AuNPs). The AuNPs/MIP/GCE showed a capability of detection and reproducibility values of 0.4 mg L-1 and 4.1%, respectively (calibration range from 0.3 to 3.1 mg L-1). The sensor was successfully applied to the determination of ethanethiol in spiked wine samples with recoveries ranging from 99% to 107%.