Four-Dimensional Computed Tomography-Guided Valve Sizing for Transcatheter Pulmonary Valve Replacement
Published on: January 20, 2022
Bespoke valve sizing avoids annular rupture in patients treated with a balloon-expandable transcatheter heart valve
Jorge Chavarria1, Felipe Falcao1, Gustavo Dutra1
1Division of Cardiology, Hamilton Health Sciences, McMaster University, Hamilton, Canada.
Background:
Annular and left ventricular outflow tract (LVOT) calcification increase the risk of annular rupture following transcatheter aortic valve replacement (TAVR). The outcomes of a strategy of routine use of a balloon-expandable valve (BEV) for all patients irrespective of annular or LVOT calcium is unknown.
Objectives:
We evaluated the impact of bespoke sizing on annular rupture in patients treated with a BEV.
Methods:
All consecutive patients undergoing TAVR at a single centre (February 2020-February 2022) were treated only with a BEV. No other valve design was used. Annular/LVOT calcification was assessed using a standardized grading system. For each annular area, we determined the percentage valve oversizing with nominal deployment. The balloon deployment volume was then adjusted when required (over-/underfilled) to achieve over-sizing of approximately 5% in the presence of annular/LVOT calcium and 5%-10% in the absence of annular/LVOT calcium. Adjusted valve areas were assumed to change proportionately to the change in balloon deployment volume.
Results:
Among 533 TAVR treated patients, annular/LVOT calcification was present in 166 (31.1%) and moderate or severe in 90 (16.9%). In patients with annular/LVOT calcification, the adjusted oversizing was 3.5 ± 3.6% and in patients without annular/LVOT calcification, the adjusted oversizing was 6.8 ± 4.7% (p < 0.001). There were no cases of annular rupture and no cases with more than mild paravalvular leak (PVL). Mild PVL was more frequent in patients with annular/LVOT calcium (10.8% vs 4.6%, p = 0.01).
Conclusion:
Bespoke BEV sizing by adjustment of balloon deployment volume avoided annular rupture in patients undergoing TAVR.
More Related Videos
05:31Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
12:17Full-root Aortic Valve Replacement by Stentless Aortic Xenografts in Patients with Small Aortic Roots
Published on: May 21, 2017
Related Concept Videos
Mitral Stenosis III: Medical Management
Aneurysm III: Interprofessional Care