Reliability and Measurement Error of Breast Volume Calculation Using Three-Dimensional Surface Imaging
Nicola Fearn1, Farid Meybodi2,3, Sharon Kilbreath1
1Sydney School of Health Sciences, Faculty of Medicine and Health, The University of Sydney, Sydney, Australia.
Lymphatic Research and Biology
|September 26, 2022
Summary
Three-dimensional surface imaging (3D-SI) reliably measures breast volume post-cancer. Vectra 3D Analysis Module (VAM) software showed lower measurement error and better reliability than Breast Sculptor for quantifying breast lymphedema.
Area of Science:
- Oncology
- Medical Imaging
- Biomedical Engineering
Background:
- Breast lymphedema quantification is challenging.
- Three-dimensional surface imaging (3D-SI) is a potential method for breast volume measurement.
- The measurement properties of 3D-SI software require thorough evaluation.
Purpose of the Study:
- To compare measurement equivalency between two 3D-SI software programs: Vectra 3D Analysis Module (VAM) and Breast Sculptor.
- To assess the reliability, standard error of measurement (SEM), and smallest detectable change (SDC) of these software programs.
- To determine the clinical utility of 3D-SI for quantifying breast lymphedema.
Main Methods:
- Retrospective analysis of 100 3D-SI scans of breasts from breast cancer patients.
- Three assessors used standardized techniques with VAM and Breast Sculptor software.
- Evaluated intra- and inter-rater reliability, SEM, and SDC for both software programs.
Main Results:
- VAM demonstrated excellent intra- and inter-rater reliability (SEM: 4.1% and 8.7%).
- Breast Sculptor showed excellent intra-rater and substantial inter-rater reliability, but with larger SEM (14.5% and 19.1%).
- Lin's concordance indicated poor agreement between the two software programs (0.81-0.88).
Conclusions:
- VAM software provides highly reliable breast volume measurements with lower error compared to Breast Sculptor.
- Significant variations in SEM and SDC exist between the software programs.
- While promising for lymphedema assessment, further improvements in 3D-SI software are needed to enhance clinical utility.