A C-arm photon counting CT prototype with volumetric coverage using multi-sweep step-and-shoot acquisitions.
Kevin Treb1, Xu Ji1, Mang Feng1
1Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, 1111 Highland Avenue, Madison, WI 53705, United States of America.
This study introduces a new C-arm photon counting detector (PCD) CT system that significantly improves image quality and low-contrast detectability over existing flat panel detector (FPD) systems. The novel step-and-shoot method enables high-quality volumetric imaging for interventional procedures.
Area of Science:
- Medical Imaging
- Radiological Technology
- Photon Counting Detectors
Background:
- Current clinical C-arm interventional systems rely on flat panel detectors (FPDs) for cone-beam CT (CBCT) imaging.
- FPD-CBCT offers volumetric coverage but lacks sufficient low-contrast detectability for specific interventional needs.
Purpose of the Study:
- To develop and evaluate a C-arm photon counting detector (PCD) CT system using a step-and-shoot acquisition method.
- To enhance tomographic imaging performance for interventional procedures.
Main Methods:
- A cadmium telluride-based PCD was integrated into a biplane C-arm system alongside an FPD.
- A synchronized motorized patient table facilitated a step-and-shoot data acquisition strategy with multiple C-arm sweeps.
- Experiments utilized physical phantoms, a human cadaver head, and an in vivo swine subject, comparing results with FPD-CT under matched conditions.
Main Results:
- PCD-CT images demonstrated lower noise and superior visualization of low-contrast lesions compared to FPD-CT.
- Improved visibility of small iodinated blood vessels and finer structures was achieved with PCD-CT.
- No geometric distortion artifacts were observed in the multi-planar PCD-CT images.
Conclusions:
- This work presents the first successful demonstration of high-quality, multi-planar PCD-CT imaging using a rotating C-arm gantry.
- The developed system shows significant potential for improving diagnostic capabilities in interventional radiology.