Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Absorbed-dose-to-water measurement using alanine in ultra-high-pulse-dose-rate electron beams
Alexandra Bourgouin1, Thomas Hackel1, Marco Marinelli2
1Physikalisch-Technische Bundesanstalt (PTB), Braunschweig, Germany.
Abstract:
Objective. The aim of the presented study is to evaluate the dose response of the PTB's secondary standard system, which is based on alanine and electron spin resonance (ESR) spectroscopy measurement, in ultra-high-pulse-dose-rate (UHPDR) electron beams.Approach. The alanine dosimeter system was evaluated in the PTB's UHPDR electron beams (20 MeV) in a range of 0.15-6.2 Gy per pulse. The relationship between the obtained absorbed dose to water per pulse and the in-beamline charge measurement of the electron pulses acquired using an integrating current transformer (ICT) was evaluated. Monte Carlo simulations were used to determine the beam quality conversion and correction factors required to perform alanine dosimetry.Main results. The beam quality conversion factor from the reference quality60Co to 20 MeV obtained by Monte Carlo simulation, 1.010(1), was found to be within the standard uncertainty of the consensus value, 1.014(5). The dose-to-water relative standard uncertainty was determined to be 0.68% in PTB's UHPDR electron beams.Significance. In this investigation, the dose-response of the PTB's alanine dosimeter system was evaluated in a range of dose per pulse between 0.15 Gy and 6.2 Gy and no evidence of dose-response dependency of the PTB's secondary standard system based on alanine was observed. The alanine/ESR system was shown to be a precise dosimetry system for evaluating absorbed dose to water in UHPDR electron beams.
More Related Videos
Related Concept Videos
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Biological Effects of Radiation
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Atomic Absorption Spectroscopy: Overview
When irradiated by EMR of a particular wavelength, these...
Atomic Emission Spectroscopy: Overview
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...