Modern Low-Field MRI of the Musculoskeletal System: Practice Considerations, Opportunities, and Challenges
Iman Khodarahmi1, Mahesh B Keerthivasan2, Inge M Brinkmann2
1From the Department of Radiology, New York University Grossman School of Medicine.
Abstract:
Magnetic resonance imaging (MRI) provides essential information for diagnosing and treating musculoskeletal disorders. Although most musculoskeletal MRI examinations are performed at 1.5 and 3.0 T, modern low-field MRI systems offer new opportunities for affordable MRI worldwide. In 2021, a 0.55 T modern low-field, whole-body MRI system with an 80-cm-wide bore was introduced for clinical use in the United States and Europe. Compared with current higher-field-strength MRI systems, the 0.55 T MRI system has a lower total ownership cost, including purchase price, installation, and maintenance. Although signal-to-noise ratios scale with field strength, modern signal transmission and receiver chains improve signal yield compared with older low-field magnetic resonance scanner generations. Advanced radiofrequency coils permit short echo spacing and overall compacter echo trains than previously possible. Deep learning-based advanced image reconstruction algorithms provide substantial improvements in perceived signal-to-noise ratios, contrast, and spatial resolution. Musculoskeletal tissue contrast evolutions behave differently at 0.55 T, which requires careful consideration when designing pulse sequences. Similar to other field strengths, parallel imaging and simultaneous multislice acquisition techniques are vital for efficient musculoskeletal MRI acquisitions. Pliable receiver coils with a more cost-effective design offer a path to more affordable surface coils and improve image quality. Whereas fat suppression is inherently more challenging at lower field strengths, chemical shift selective fat suppression is reliable and homogeneous with modern low-field MRI technology. Dixon-based gradient echo pulse sequences provide efficient and reliable multicontrast options, including postcontrast MRI. Metal artifact reduction MRI benefits substantially from the lower field strength, including slice encoding for metal artifact correction for effective metal artifact reduction of high-susceptibility metallic implants. Wide-bore scanner designs offer exciting opportunities for interventional MRI. This review provides an overview of the economical aspects, signal and image quality considerations, technological components and coils, musculoskeletal tissue relaxation times, and image contrast of modern low-field MRI and discusses the mainstream and new applications, challenges, and opportunities of musculoskeletal MRI.
Related Concept Videos
Magnetic Resonance Imaging
Microscopic Anatomy of Skeletal Muscles
The muscle sarcolemma is a plasma membrane enclosing each muscle cell that conducts electrical signals called action potentials. The sarcolemma extends into the cell to form T-tubules, ensuring the neural impulses are uniformly distributed across the entire muscle...
Gross Anatomy of Skeletal Muscles
Imaging Studies IV: Magnetic Resonance Imaging
Gross Anatomy of Bone
The diaphysis is the tubular shaft that runs between the proximal and distal ends of the bone. The walls of the diaphysis are composed of dense and hard compact bone made of numerous osteons — the functional unit of the compact bone. The hollow region in the diaphysis is called the medullary cavity, which harbors the bone marrow. In infants and children, this marrow cavity is filled with red marrow, whereas in...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...