Analysis of water quality over non-condensable gases concentration on steam used for sterilization
Emerson Aparecido Miguel1, Paulo Roberto Laranjeira2, Marina Ishii3
1Orionce Serviços de Metrologia Ltda., Barueri, SP, Brazil.
Background:
Non-condensable gases (NCGs) are all gases that do not undergo liquefaction during the saturated steam sterilization process. During a sterilization cycle, the NCGs presence inside the chamber is one of the biggest threats to the sterilization process compromising process validation and product quality.
Methods:
In this work, 170 testing of NCGs concentrations performed between September 2016 and August 2021 were carried out by Orionce Serviços de Metrologia Ltda (Barueri, SP, Brazil), according to EN285:2015 procedure. For steam generation, the types of water used were softened water (SW), one-step reverse osmosis (RO), purified water (PW), and water for injection (WFI). The data obtained were analyzed using Minitab® software, version 18.1, to identify a relationship between the water quality used for steam generation and the concentration of non-condensable gases determined into equipment qualification.
Findings:
From total tests performed, 109 tests passed, and 61 tests failed, representing 64.2% and 35.8%, respectively. A higher failure rate was observed in terms of concentration of non-condensable gases in systems that used purified water for steam generation (64.7%), followed by softened water (55.6%), one-step reverse osmosis (42.9%), and water for injection (7.6%).
Conclusion:
System processes using WFI for steam generation showed better results for steam quality approvement, in terms of NCGs concentration, compared to softened, purified, or reverse osmosis water treatments in the concentration of non-condensable gases in steam used for sterilization processes of industrial utilities.
Related Concept Videos
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Quality of Water
Physical Methods for Controlling Microbial Growth: Temperature
Volatilization
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...