Comparative efficacy of finger versus forehead Plethysmographic Variability Index monitoring in pediatric surgical
Seyfullah Fidan1, Seniyye Ülgen Zengin1, Meliha Orhon1
1Department of Anesthesiology and Reanimation, Marmara University School of Medicine, Istanbul, Turkey.
Introduction:
The Plethysmographic Variability Index can be measured by both finger and forehead probes. Vasoconstriction may jeopardize the reliability of finger PVI measurements in pediatric patients undergoing surgery. However, forehead vasculature exhibits more marked resistance to alterations in the vasomotor tonus.
Objective:
Our aim was to compare the Plethysmographic Variability Index measured via finger or forehead probes in mechanically ventilated pediatric surgery patients in terms of their ability to predict fluid responsiveness as well as to determine the best cut-off values for these two measurements.
Materials And Methods:
A total of 50 pediatric patients undergoing minor elective surgery were included after provision of parental consent and ethics committee approval. Perfusion index measured at the finger or forehead and Plethysmographic Variability Index monitoring comprised the primary assessments. Hemodynamic parameters monitored included perfusion index, Plethysmographic Variability Index, and cardiac output. A ≥ 15% increase in cardiac output following passive leg raise maneuver was considered to show fluid responsiveness. Two groups were defined based on fluid responsiveness: Group R (responsive) and Group NR (non-responsive). Student's t-test, Mann-Whitney U test, DeLong test, and ROC were used for statistical analysis.
Results:
The area under curve for finger and forehead Plethysmographic Variability Index prior to passive leg raise maneuver were 0.699 (p = .011) and 0.847 (p < .001), respectively. The sensitivity for finger and forehead measurements at a cut-off value of ≤14% was 92.9% and 96.4%, and 45.4% and 72.7%, respectively.
Conclusion:
Although forehead and finger Plethysmographic Variability Index monitoring were similarly sensitive in predicting fluid responsiveness in pediatric surgical patients, the former method provided higher specificity. The best cut-off value for PVI measurements with forehead and finger probes was found to be 14%.
More Related Videos
05:01A Detailed Protocol for Physiological Parameters Acquisition and Analysis in Neurosurgical Critical Patients
Published on: October 17, 2017
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...