Comparison of subjective image analysis and effective dose between low-dose cone-beam computed tomography machines
Arnon Charuakkra1, Phattaranant Mahasantipiya1, Antti Lehtinen2
1Division of Oral and Maxillofacial Radiology, Department of Oral Biology and Diagnostic Sciences, Faculty of Dentistry, Chiang Mai University, Chiang Mai, Thailand.
Objectives:
To compare the cone-beam computed tomography (CBCT) image quality and effective dose between low-dose scanning and standard manufacturer-recommended protocols among different CBCT units.
Methods:
Three human-equivalent phantoms were scanned using the ultra-low-dose (ULD), low dose (LD), and standard dose (STD) modes of ProMax 3D Mid (Planmeca Oy, Helsinki, Finland) and Orthophos SL (Sirona, Bensheim, German) for the CBCT images. The quality of the dental anatomical images was assessed by four experienced oral and maxillofacial radiologists using a 5-point Likert scale. OnDemand3D (Cybermed Co., Seoul, Korea) was used as the third-party software for viewing. The percentage of absolute agreement was calculated to determine intra- and interrater agreements among the observers. The effective doses for all CBCT scanning protocols were also calculated.
Results:
The STD protocol yielded a higher image quality than did the ULD and LD protocols in both ProMax 3D Mid and Orthophos SL. The ULD and LD protocols demonstrated an "acceptable-to-good" sense of visual perception of the CBCT images. The visibility scores significantly differed between the ULD and LD and the STD protocols in ProMax 3D Mid and Orthophos SL, except for the 120-kVp setting in ProMax 3D Mid. The average intra- and interrater agreement scores ranged from 0.63 to 0.89 and from 0.44 to 0.76, respectively. The ULD and LD protocols reduced the radiation dose sixfold compared with the STD protocol.
Conclusions:
High-tube-voltage protocols could remarkably reduce the imaging dose without degrading the image quality. Specifically, ULD and LD CBCT protocols may be adopted as routine practice for diagnosis and treatment planning.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Imaging
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT