Background:
Invasive assessments of microvascular function are rapidly becoming an integral part of physiological assessment in chronic coronary syndromes.
Objective:
We aimed to establish a reference range for Doppler flow velocity-derived hyperaemic microvascular resistance index (HMR) in a cohort of angina with no significant epicardial coronary obstruction (ANOCA) patients with no structural pathophysiological alterations in the coronary circulation.
Methods:
The reference population consisted of ANOCA patients undergoing invasive coronary vasomotor function assessment who had a coronary flow reserve (CFR) >2.5, and had either (1) tested negatively for spasm provocation (n = 12) or (2) tested positively with only angina at rest (n = 29). A reference range for HMR was established using a non-parametric method and correlations with clinical characteristics were determined using a spearman rank correlation analysis.
Results:
In 41 patients median HMR amounted to 1.6 mmHg/cm/s [Q1, Q3: 1.3, 2.2 mmHg/cm/s]. The reference range for HMR that is applicable to 95% of the population was 0.8 mmHg/cm/s (90% CI: 0.8-1.0 mmHg/cm/s) to 2.7 mmHg/cm/s (90% CI: 2.6-2.7 mmHg/cm/s). No significant correlations were found between HMR and clinical characteristics.
Conclusion:
In this reference population undergoing invasive coronary vasomotor function testing, the 90% confidence interval of the HMR upper limit of normal ranges from 2.6 to 2.7 mmHg/cm/s. A > 2.5 mmHg/cm/s HMR threshold can be used to identify abnormal microvascular resistance in daily clinical practice.