Evaluation of temperature of a full ceramic total knee arthroplasty during MRI examinations
Klemens Trieb1,2, Andreas Artmann3, Michael Krupa3
1Department of Orthopaedic and Trauma Surgery, Paracelsus Medical University Salzburg, Salzburg, Austria.
Background:
A diagnosis by magnetic resonance imaging (MRI) is often necessary before surgery of degenerative spine diseases. This can lead to a possible conflict with an inserted implant of the hip or knee. Heat generation or movement could be caused by the magnetic field. The aim of this study is to investigate temperature development in vitro in a 1.5T MRI of a ceramic knee arthroplasty.
Methods:
A full ceramic, complete metal-free non-constrained primary total knee arthroplasty is investigated. Temperature change was measured between platinum resistors before and after each MRI sequence by change of resistance. The knee implant was placed in a plastic container after the sensors were attached. Then the container was completely filled with ultrasound gel. To document any possible movement of the implant, a grid was placed under the container to document the position of the implant before and after the scans.
Results:
A total of four standard knee sequences were performed. The temperature at sites 1 to 5 per implant was always documented in the as-is state before MRI and then after each sequence. A total of 5 temperature measurement points were taken per implant. It was found that there were extremely small temperature variations. These were always in the range of less than 1°C. There was no case of movement of the implant triggered by the MRI scan.
Conclusions:
The experimental investigations carried out here showed homogeneous results with this experimental setup. It is concluded that, at least in vitro, that this ceramic knee implant can be used in MRI examinations without heating or movement.
More Related Videos
Related Concept Videos
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Imaging Studies IV: Magnetic Resonance Imaging
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...