On the occurrence of very low intra-particle diffusion rates in zwitterionic hydrophilic interaction liquid
Haibin Li1, Gert Desmet2, Zhengjin Jiang3
1University of Leuven (KU Leuven), Department for Pharmaceutical and Pharmacological Sciences, Pharmaceutical Analysis, Herestraat 49, Leuven, Belgium; Institute of Pharmaceutical Analysis, College of Pharmacy, Jinan University, Guangzhou 510632, China.
Abstract:
Intra-particle diffusion is investigated in particle-packed and monolithic zwitterionic hydrophilic interaction liquid chromatography (HILIC) polymer columns. For this purpose, plate heights are first measured over a broad range of velocities for polar compounds with zone retention factors between k″ = 2 and 10 on all column types, multiplied with the corresponding velocity, and extrapolated to zero velocity. The thus obtained B-term coefficients are additionally verified via peak parking experiments and indicate a very low degree of longitudinal diffusion in all columns, that moreover seems to be independent of the zone retention factor. This is in contrast to earlier observations on reversed-phase liquid chromatography columns and bare silica HILIC columns. To obtain more insight into the reasons for these low levels of longitudinal diffusion, the experimentally obtained effective diffusion coefficients are modeled to equations derived from the effective medium theory, such as the models developed by Torquato and Jefrey & Hashin. It is, however, demonstrated that these models fail to adequately describe effective diffusion in the studied zwitterionic HILIC columns, due to the low degree of intraparticle diffusion in these stationary phases. Fitting the experimental effective diffusion coefficients to the two limiting cases of the effective medium theory, i.e., the pure parallel-connection and pure serial-connection case, it is observed that to stay within these limits, intraparticle diffusion must be extremely low, i.e., Dpz/Dm ≤ 0.01, indicating that the diffusion inside the pores is at least 100 times smaller than the bulk diffusion coefficient. There are several possible explanations for these low intraparticle diffusion coefficients, such as a strongly hindered diffusion in the polymer matrix, extremely low surface diffusion rates in the aqueous layer adsorbed onto the stationary phase and/or the occurrence of slow localized adsorption events.
Related Concept Videos
Diffusion on Chromatography Columns
Longitudinal diffusion occurs when the solute molecules in the mobile phase diffuse from the more concentrated center of the chromatographic band to the more dilute regions on either side, both towards and against the flow direction. This...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Column Efficiency: Rate Theory
During elution, a solute molecule experiences numerous transitions between stationary and mobile phases, exhibiting irregular residence times in...
Analyte Adsorption and Distribution