Related Experiment Video
Updated: Aug 27, 2025

Fabrication and Operation of an Oxygen Insert for Adherent Cellular Cultures
Published on: January 6, 2010
Evaluation of an innovative low flow oxygen blender system for global access
Ellie Ng1, Michelle Dundek1,2, Thomas F Burke1,2,3,4
1Vayu Global Health Foundation, Boston, MA, United States.
Background:
Safe and effective oxygen delivery methods are not available for the majority of infants and young children globally. A novel oxygen blender system was designed to accurately deliver concentration-controlled, oxygen-enriched air to hypoxemic children up to age five. The system does not require compressed medical air, is compatible with both oxygen tanks and oxygen concentrators, and is low cost. This is the first study that tested the performance of the innovative oxygen blender system.
Methods:
The performance of the oxygen blender system was assessed in vitro based on delivered oxygen levels and flow rates with an oxygen tank, an oxygen tank using a nasal occlusion model, and an oxygen concentrator.
Results:
The measured %O2 of the performance test was within ± 5% of full scale (FS) of the target value across all flows and all nasal cannulas. Occlusion testing demonstrated that 50% occlusion did not significantly affect the system outputs. The oxygen blender system was shown to be compatible with both oxygen tanks and oxygen concentrators.
Conclusions:
The novel oxygen blender system accurately controls oxygen concentrations and blended air flow rates, and is compatible with both oxygen tanks and oxygen concentrators. This innovation may be an opportunity for improved infant and child oxygen treatment worldwide.
More Related Videos
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Treatment for Pulmonary Arterial Hypertension: Oxygen Therapy for Respiratory Failure
Oxygen therapy is vital in increasing and maintaining blood oxygen levels in PAH patients. As a result, it aids in reducing fatigue,...