Performance comparison of two nucleic acid amplification systems for SARS-CoV-2 detection: A multi-center study
Chan Mo1, Kamfai Lo2, Ying He3
1Medical Laboratory of the Third Affiliated Hospital of Shenzhen University, Shenzhen, Guangdong, China.
Journal of Clinical Laboratory Analysis
|October 5, 2022
Summary
Rapid nucleic acid testing for COVID-19 shows high accuracy, especially for high viral loads. This rapid test is a valuable tool for identifying SARS-CoV-2 infections and complements existing detection methods.
Area of Science:
- Molecular Diagnostics
- Infectious Disease Testing
- COVID-19 Detection
Background:
- Emergence of rapid nucleic acid testing systems to combat COVID-19.
- Limited comparative studies on diagnostic performance versus conventional systems.
- Need for evaluating clinical utility of rapid SARS-CoV-2 detection.
Purpose of the Study:
- Evaluate and compare the clinical diagnostic performance of a rapid nucleic acid detection system against conventional methods for SARS-CoV-2.
- Assess the accuracy, including sensitivity, specificity, and predictive values.
- Analyze the concordance and performance based on viral load.
Main Methods:
- Retrospective analysis of 63,352 oropharyngeal swab results.
- Comparison of sensitivity, specificity, PPV, and NPV between rapid (Daan Gene) and conventional systems.
- Assessment of concordance using OPA, PPA, NPA, and kappa (κ) value.
- Analysis of rapid system sensitivity relative to viral load (Ct values).
Main Results:
- The Daan Gene rapid system demonstrated slightly lower sensitivity (0.86) but equivalent specificity compared to conventional systems (0.979).
- High positive predictive value (PPV) and negative predictive value (NPV) of ≥98.0% for SARS-CoV-2.
- Excellent agreement between Daan Gene and conventional systems (κ = 0.893).
- Daan Gene achieved 99.31% sensitivity for high viral load (Ct < 35) and 50% for low viral load (Ct ≥ 35).
Conclusions:
- Rapid nucleic acid detection systems offer a viable alternative for rapid SARS-CoV-2 identification, particularly in cases with high viral loads.
- The Daan Gene system shows excellent concordance with conventional methods, indicating reliable performance.
- This rapid testing approach can serve as a valuable complement to existing COVID-19 diagnostic strategies.