Dual-Applicator MR Imaging-Guided Microwave Ablation with Real-Time MR Thermometry: Phantom and Porcine Tissue Model
Liqiang Ren1, David A Woodrum1, Krzysztof R Gorny1
1Department of Radiology, Mayo Clinic, Rochester, Minnesota.
Dual microwave ablation (MWA) applicators create larger treatment zones during MR imaging-guided procedures. Despite a slight decrease in MR thermometry image quality, dual applicators allow adequate monitoring of ablation zones.
Area of Science:
- Medical Imaging
- Interventional Radiology
- Biomedical Engineering
Background:
- Magnetic resonance (MR) imaging-guided microwave ablation (MWA) is an effective cancer treatment.
- Optimizing applicator configuration is crucial for treatment efficacy and safety.
- Real-time MR thermometry provides essential feedback during MWA procedures.
Purpose of the Study:
- To evaluate the impact of simultaneous dual applicator use on MR thermometry image quality.
- To characterize the treatment zone patterns generated by dual applicators during MR-guided MWA.
- To assess the feasibility of dual-applicator MWA for creating larger ablation zones.
Main Methods:
- MWA experiments were conducted on a 1.5-T MR scanner using commercial microwave systems.
- Phantom studies assessed the effect of dual applicators on MR image quality.
- Porcine tissue models were used with real-time MR thermometry, comparing single and dual applicators at varying distances (2.6, 3.6, 4.6 cm).
- Fiberoptic probes measured surface temperature changes.
Main Results:
- Dual applicator use reduced MR thermometry signal-to-noise ratio (SNR) to 55% (vs. 86.2% for single applicator), but image quality remained adequate for visualization.
- Temperature increases at the tissue surface were not significantly different between single (18.2°C) and dual (19.7°C) applicators.
- Dual-applicator MWA produced significantly larger treatment zones (29.4–42.6 cm²) compared to single-applicator MWA (18.6 cm²).
Conclusions:
- MR imaging-guided dual-applicator MWA effectively creates larger ablation zones.
- Adequate real-time MR thermometry image quality is maintained for monitoring treatment zone evolution.
- Dual applicator MWA offers a promising approach for enhanced thermal ablation therapies.