Optimisation of imaging parameters during CT guided interventional procedures in patients with total hip arthroplasty
Anish Patel1, Christine Azzopardi1, Scott Evans2
1Department of Musculoskeletal Imaging, Royal Orthopaedic Hospital, Birmingham, UK.
Objective:
To determine the optimal KV and mAs settings during CT guided injections in the presence of a total hip arthroplasty so that image quality is maintained whist keeping the dose as low as reasonably achievable.
Methods:
A total hip arthroplasty phantom with needles with differing gauges was scanned using different CT parameters (from low dose to high dose) and evaluating if this had any effect on needle conspicuity. Conspicuity was graded from 1 to 3 by 2 independent blinded reviewers.
Results:
Irrespective of the CT settings used (high dose or low dose parameters) needle conspicuity was not adversely affected by the THA for either scorer, therefore a kVp of 100 mA and a. In addition the needle gauge did not affect the conspicuity of the needle.
Conclusion:
CT guided injections in this total hip arthroplasty phantom model can be performed without any adverse effect on the conspicuity of the needle tip on low dose CT settings.
Advancement In Knowledge:
This paper enables one to optimise the kV and mA while performing interventional procedures.
More Related Videos
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
05:37Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Imaging Studies VII: Vascular Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...