Related Experiment Video
Updated: Aug 26, 2025

A Novel Method to Determine the Longitudinal Antibacterial Activity of Drug-Eluting Materials
Published on: March 3, 2023
Investigation and analysis of four countries' recalls of osteosynthesis implants and joint replacement implants from
Yang Wang1,2, Kai Xu1, Yuchen Wang1
1Department of Orthopedic Surgery, The First Affiliated Hospital of Soochow University, No. 899 Ping Hai Road, Suzhou, 215000, China.
Background:
Medical devices are used in almost all orthopedic surgical subspecialties, and the frequency of adverse events is increasing, which should not be ignored. To provide suggestions on how to avoid implant recalls from the perspective of manufacturers, medical institutions and supervisions, as well as how to respond promptly to adverse events.
Methods:
The research extracted recalls of osteosynthesis implants and joint replacement implants from January 1, 2011, to June 30, 2021, in the CNMPA, FDA, HC and ATGA websites and collected the information on device name, recall time, recall class, recall manufacturer, device classification and affected areas. Moreover, the McKinsey 7S model and fishbone diagram were used to analyze recall reasons.
Results:
A total of 315 cases of osteosynthesis implants and 286 cases of joint replacement implants were reported in China, the USA, Canada and Australia. The recalls number from 2016 to 2021 was more than that from 2011 to 2015 for osteosynthesis implant (p = 0.012) and joint replacement implant (p = 0.002), and both mainly focused on class II (76.19% and 78.32%). There were statistical differences in the four countries for both implants (p = 0.000), especially osteosynthesis implant between China and the USA (p = 0.000), China and Canada (p = 0.001), the USA and Australia (p = 0.002), and joint replacement implant between China and Australia (p = 0.000).
Conclusions:
To avoid the recalls of such implants, manufacturers should strictly select implant materials and components, develop detailed labels and instructions, severely control the packaging process and establish the integrity of medical device data. Medical institutions should standardize procurement procedures, use qualified equipment and train medical workers. It also requires supervisions to conduct premarket safety assessments. In addition, regulators should strengthen supervision and establish reporting systems to deal with the occurrence of adverse events promptly.