Related Experiment Video
Updated: Aug 26, 2025

Quantification of the Immunosuppressant Tacrolimus on Dried Blood Spots Using LC-MS/MS
Published on: November 8, 2015
Application of a new volumetric microsampling device for quantitative bioanalysis of immunosuppression
Abbie D Leino1, John Takyi-Williams2, Bo Wen2
1Department of Clinical Pharmacy, College of Pharmacy, University of Michigan, Ann Arbor, MI 48109, USA.
Abstract:
Background: Volumetric absorptive microsampling may reduce the blood collection burden associated with therapeutic drug monitoring of immunosuppression to prevent organ transplant rejection. This work describes the development of a laboratory and analytical technique for quantifying tacrolimus and mycophenolic acid (MPA) from the Tasso-M20™ in human whole blood using bead-based impact-assisted extraction. Results: The sampled blood volume was accurate with estimated volumes within <2% of the expected 20 μl. Recovery using impact-assisted extraction was 73-87% for MPA and 100% for tacrolimus and was hematocrit-independent for both analytes. The LC-MS/MS assay is precise and accurate within the acceptance criteria of 15%. Conclusion: The sampling and extraction procedures allowed for accurate quantification of tacrolimus and MPA. Exploration of abuse scenarios identified important education points for patients conducting home-based sample collections in the future.