Inter-transducer variability of ultrasound image quality in obese adults: Qualitative and quantitative comparisons
Megan Maar1, Juhyun Lee2, Anthony Tardi3
1Rocky Vista University, Ivins, UT, USA.
Purpose:
Acquiring high-quality ultrasound images of deep abdominal organs and vasculatures in obese adults (BMI >30 kg/cm2) is considered challenging. The aim of the study was to assess the inter-transducer variability in B-mode and color Doppler image quality from four commercial ultrasound transducers through qualitative and quantitative analyses.
Methods:
Four curvilinear transducers on three ultrasound scanners were used to acquire B-mode and color Doppler images of deep abdominal structures in 15 obesity ≥ class II (BMI >35 kg/cm2) adults. Using visual-qualitative assessment and an offline image processing software, visual-qualitative score and quantitative mean pixel values of B-mode images, and color area ratios of color Doppler images were calculated. Differences in these values among the transducers were analyzed using one-way ANOVA. The intra- and inter-observer reliability of visual-qualitative assessment and offline image processing was tested using the intraclass correlation coefficient (ICC).
Results:
Differences in visual-qualitative score, mean pixel value of B-mode images, and color area ratio of color Doppler images among the four transducers were significant (p < 0.001). Transducer -4 produced the highest quality of B-mode (45-53% improvement) and color Doppler (22-73% improvement) images among the transducers. Intra-observer repeatability and inter-observer reproducibility were higher with performing offline image processing than visual-qualitative assessment (ICC: 0.97-0.99 versus ICC: 0.76-0.97).
Conclusion:
There was significant image quality variability between different transducers. Transducer -4, a transducer designed specifically for high BMI patients, had the highest quality B-mode and color Doppler images compared to the other transducers lending to improved ultrasonographic visualization in obese patients.
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Imaging Studies II: Ultrasonography
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Imaging Studies III: Computed Tomography