The Effect of Tedlar Bags on the Composition of Exhaled Human Breath Samples
Jiuyan Zhao1, Lanlan Zhu2, Wei Zhang2
1Emergency Department, Guizhou Province, The Affiliated Hospital of Guizhou Medical University, No. 28, Guiyijie Road, Liuguangmen, Guiyang City 550004, China.
Abstract:
The PVF Tedlar is widely used for gas collection in clinical diagnostics and environmental research. However, sample collection is frequently associated with the degradation, adsorption, or transformation of sensitive chemicals. Here, we explore to what extent the Tedlar bag collection effects the composition of expired breath samples. Collected breath samples were analyzed using the EESI-MS technique after the storage time of 30 min, 1 h, 2 h, 3 h, 4 h, 5 h, and 6 h, respectively. Our results demonstrated the gradual MS signal decay after 3 h storage. The decay rate of 3 h is about 45% and 6 h is about 88%. Therefore, the Tedlar bag is suggested as a reliable breath holder on the time scale of <3 h.