An Automation System Equivalent to the Douglas Bag Technique Enables Continuous and Repeat Metabolic Measurements in
Koichiro Shinozaki1, Pey-Jen Yu2, Qiuping Zhou3
1The Feinstein Institutes for Medical Research, Northwell Health, Manhasset, New York; Department of Emergency Medicine, Zucker School of Medicine at Hofstra/Northwell, New York, New York; Department of Emergency Medicine, North Shore University Hospital, Manhasset, New York.
Purpose:
To develop a system that is equivalent to the gold standard Douglas Bag (DB) technique for measuring oxygen consumption (V̇o2), carbon dioxide generation (V̇co2), and respiratory quotient (RQ) and to validate its use in clinical settings.
Methods:
This was a prospective, observational study conducted at a suburban, quaternary care teaching hospital. Healthy volunteers and patients 18 years or older who received mechanical ventilation were enrolled.
Findings:
Data from 3 healthy volunteers and 7 patients were analyzed in this study. The interrater reliability between the automation device and DB methods were 0.999, 0.993, and 0.993 for V̇o2, V̇co2, and RQ, respectively. In healthy volunteers, mean (SD) V̇o2, V̇co2, and RQ measured by DB were 411 (100) mL/min, 288 (79) mL/min, and 0.70 (0.03) at high fraction of inspired oxygen (Fio2) and 323 (46) mL/min, 280 (45) mL/min, and 0.85 (0.05) at normal Fio2, respectively. V̇o2 was significantly higher (P < 0.05) and RQ was lower (P < 0.01) in the high Fio2 group as compared to those in the normal Fio2 group. Values measured by the automation system were 227 (31) mL/min, 141 (18) mL/min, and 0.62 (0.04) at high Fio2 and 209 (25) mL/min, 147 (18) mL/min, and 0.70 (0.06) at normal Fio2, respectively. RQ was significantly lower (P < 0.05) in the high Fio2 group as compared to the normal Fio2 group. We also successfully performed continuous and repeat measurements by using the device. The longest measurement reached 12 hours 15 minutes, including 50 cycles of repeat measurements that are equivalent to the DB technique as described above.
Implications:
We developed an automation system that enables repeat measurements of V̇o2, V̇co2, and RQ, and the accuracy was equivalent to the DB technique. High Fio2 may decrease RQ because of an increase in V̇o2.
More Related Videos
13:10Evaluation of Respiratory System Mechanics in Mice using the Forced Oscillation Technique
Published on: May 15, 2013
07:09Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
Published on: February 20, 2017
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation I: Indication and Settings
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation: