Monitoring the Awake and Anesthetized Unconscious States Using Bispectral Index and Electroencephalographic
Marianne Cecilie Johansen Nævra1, Luis Romundstad2, Anders Aasheim3
1Section of Clinical Neurophysiology, Department of Neurosurgery, Division of Clinical Neuroscience, Rikshospitalet, Oslo University Hospital, Oslo, Norway.
Abstract:
Objective. Our objective was to compare three electroencephalography (EEG)-based methods with anesthesiologist clinical judgment of the awake and anesthetized unconscious states. Methods. EEG recorded from 25 channels and from four channel bilateral Bispectral index (BIS) electrodes were collected from 20 patients undergoing surgery with general anesthesia. To measure connectivity we applied Directed Transfer Function (DTF) in eight channels of the EEG, and extracted data from BIS over the same time segments. Shannon's entropy was applied to assess the complexity of the EEG signal. Discriminant analysis was used to evaluate the data in relation to clinical judgment. Results. Assessing anesthetic state relative clinical judgment, the bilateral BIS gave the highest accuracy (ACC) (95.4%) and lowest false positive discovery rate (FDR) (0.5%) . Equivalent DTF gave 94.5% for ACC and 2.6% for FDR. Combining all methods gave ACC = 94.9% and FDR = 1%. Generally, entropy scored lower on ACC and higher on FDR than the other methods (ACC 90.87% and FDR 4.6%). BIS showed at least a one minute delay in 18 of the 20 patients. Conclusions. Our results show that BIS and DTF both have a high ACC and low FDR. Because of time delays in BIS values, we recommend combining the two methods.
More Related Videos
10:19Induction of an Isoelectric Brain State to Investigate the Impact of Endogenous Synaptic Activity on Neuronal Excitability In Vivo
Published on: March 31, 2016
14:52Recording Brain Electromagnetic Activity During the Administration of the Gaseous Anesthetic Agents Xenon and Nitrous Oxide in Healthy Volunteers
Published on: January 13, 2018