Advancing a Bulky Delivery System Using a Novel Technique During Percutaneous Pulmonary Valve Implantation
Wenzhi Pan1, Shiqiang Hou1, Shasha Chen1
1Department of Cardiology, Zhongshan Hospital, Fudan University, Shanghai Institute of Cardiovascular Disease, Shanghai, People's Republic of China.
Abstract:
During percutaneous pulmonary valve implantation, it is challenging to advance the delivery system loaded with a stent-valve through the right ventricular outflow tract. An extra-stiff wire can provide enough supporting and stable positioning to prevent the displacement of the guide wire in stent-valve delivery. In the 2 cases reported here, we used a snare to trap the extra-stiff wire and a large sheath to support the snare; thus, the extra-stiff wire was pulled to the distal end and provided strong support for the advancement of the delivery system. After 8 months of follow-up, the patients were in good condition with excellent valve function. We named this novel, efficient technique the "parallel anchor sheath snare" technique. This report highlights a novel delivery strategy for patients with tortuous right ventricular outflow tracts in percutaneous pulmonary valve implantation.