Effect of ElectroMagnetic interference from SmartPHone on cardiac ImplaNtable electronic device (EMI-PHONE study).
Sanatcha Apakuppakul1, Nilubon Methachittiphan1, Sirin Apiyasawat1
1Ramathibodi hospital Mahidol University Bangkok Thailand.
Modern smartphones do not interfere with cardiac implantable electronic devices (CIEDs). This study found no electromagnetic interference (EMI) or adverse events, indicating safe usage of smartphones with CIEDs.
Area of Science:
- Biomedical Engineering
- Electrophysiology
- Medical Device Technology
Background:
- Smartphones emit electromagnetic waves that can potentially interfere with cardiac implantable electronic devices (CIEDs).
- Previous evidence suggests electromagnetic interference (EMI) from mobile phones to CIEDs.
- Both smartphones and CIEDs have undergone design improvements to mitigate EMI.
Purpose of the Study:
- To evaluate the electromagnetic interference (EMI) potential of contemporary smartphones on cardiac implantable electronic devices (CIEDs).
- To assess the safety of using smartphones in close proximity to patients with CIEDs.
Main Methods:
- 80 subjects with CIEDs were tested with three smartphone models (Nokia 3310, iPhone 7, Samsung 9S).
- Phones were placed at 0 cm at the generator site, atrial lead, and ventricular lead levels.
- Real-time interrogations and post-test assessments were conducted during various phone usage scenarios to detect EMI and device parameter changes.
Main Results:
- No electromagnetic interference (EMI) was detected from any of the tested smartphones across all test conditions.
- No adverse events, including pacemaker inhibition, inappropriate ICD shocks, or CIED device malfunction, were observed in any participant.
- All 80 subjects completed the testing protocol without any complications.
Conclusions:
- Current generation smartphones exhibit no significant electromagnetic interference (EMI) with cardiac implantable electronic devices (CIEDs).
- Patients with CIEDs can safely use modern smartphones with reduced concern for device malfunction or adverse events.
- The findings support the safe co-existence of smartphones and CIEDs in clinical practice.
More Related Videos
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
07:41Simultaneous Electrical and Mechanical Stimulation to Enhance Cells' Cardiomyogenic Potential
Published on: January 18, 2019
Related Concept Videos
Mechanism of Cardiac Arrhythmias
Bacterial Signaling