Operator's Influence on Contrast Agent Intravasation During Hysterosalpingo-Contrast Sonography: Explanation Based on
Bin-Bin Jin1, Yan Ma1, Xiu-Hua Zhao1
1Department of Medical Ultrasound, The First Affiliated Hospital of Guangxi Medical University, Nanning, People's Republic of China.
Objective:
To analyze the risk factors of the operators on contrast agent intravasation during hysterosalpingo-contrast sonography (HyCoSy).
Methods:
We retrospectively collected 399 infertile women who underwent HyCoSy by the same sonographer. These patients were divided into two groups according to the way how the assistants connected the syringe to the uterus radiography catheter to inject the contrast agent. We analyzed whether the use of different contrast bolus injection methods had any influence on the incidence of intravasation during HyCoSy.
Results:
There was no significant difference between the two groups with different cross-sectional areas of the syringe outflow tract in the risk variables for intravasation, but the intravasation rates of the two groups were different, 26.4% in group A and 17.1% in group B, P <0.05. The primary manifestation was that when both fallopian tubes were unobstructed, the intravasation rate of group B with smaller cross-sectional area of the outflow tract of the syringe was lower, and the difference was statistically significant. The inferences drawn from our physics model were also in line with the clinical results.
Conclusion:
The influence of different operators on the contrast agent intravasation rate of HyCoSy cannot be ignored. The assistants of HyCoSy examination should inject the contrast agent slowly and steady, and a needle can be used as a flow restrictor to control the flow into the uterine cavity per unit time, slow down the rising speed of intrauterine pressure, and avoid the accumulation of contrast agent in the uterine cavity, so as to reduce the intravasation caused by operator factors.
More Related Videos
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Physiological Pharmacokinetic Models: Blood Flow-Limited Versus Diffusion-Limited Models
Imaging Studies II: Ultrasonography
Imaging Studies III: Computed Tomography
Compartment Models: Two-Compartment Model
Compartment Models: Single-Compartment Model