Related Experiment Video
Updated: Aug 25, 2025

The Portable Chemical Sterilizer PCS, D-FENS, and D-FEND ALL: Novel Chlorine Dioxide Decontamination Technologies for the Military
Published on: June 29, 2014
Splash generation and droplet dispersal in a well-designed, centralized high-level disinfection unit
Cori L Ofstead1, Krystina M Hopkins1, Frank E Daniels2
1Ofstead & Associates, Inc., St. Paul, MN.
Background:
Sterile processing personnel routinely decontaminate medical devices that are heavily soiled with blood, tissue, and secretions. Contamination may spread throughout processing areas, potentially exposing personnel and patient-ready devices, especially when there is insufficient separation between the dirty and clean areas.
Objective:
This study aimed to identify activities that generate splash, determine how far droplets travel during manual cleaning, characterize the impact of practices on splash generation, and assess effectiveness of personal protective equipment (PPE) at preventing splash exposure to technicians and visitors in the decontamination unit.
Methods:
Moisture-detection paper was affixed to PPE and environmental surfaces in a new processing department designed to optimize workflow and prevent cross-contamination. Droplet generation and dispersal were assessed during manual cleaning of a colonoscope and a transvaginal ultrasound probe.
Results:
Splash was generated by most activities and droplets were detected up to 7.25 feet away. Transporting wet endoscopes dispersed droplets on a 15-foot path from the sink to the automated endoscope reprocessor. Extensive droplets were detected on PPE worn by technicians at the sink and observers 3-4 feet away.
Conclusions:
Manual cleaning of devices generated substantial splash, drenching technicians and the environment with droplets that traveled more than 7 feet. Engineering controls and better PPE are needed to reduce personnel exposure and risks associated with the potential dispersal of contaminated fluids throughout the facility.
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Hand hygiene
Hand washing...
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Chemical Agents for Microbial Control