Related Experiment Video
Updated: Aug 25, 2025

Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
Unique challenges posed by a dysfunctional native right ventricular outflow tract for percutaneous pulmonary valve
Kothandam Sivakumar1, Ramyashri Chandrasekaran1, Ziyad M Hijazi2
1Department of Pediatric Cardiology, Institute of Cardiovascular Diseases, Madras Medical Mission, Chennai, Tamil Nadu, India.
Abstract:
Dysfunctional right ventricular outflow tracts after a repair for tetralogy of Fallot using a transannular patch offer limited nonsurgical opportunities due to their large dimensions. A discrete subannular narrowing between a dilated right ventricle and the enlarged pulmonary trunk was a potential anatomical target for the creation of a landing zone using a prestent in a young male with severe pulmonary regurgitation and moderate stenosis. Asymmetric expansion of the prestent in the angulated outflow tract led to distal stent embolization that was stabilized by another telescoping stent before successful valve implantation. This manuscript details the unique challenges posed by a dilated regurgitant outflow tract for implanting a pulmonary valve.