Advanced Vision Analyzer-Virtual Reality Perimeter: Device Validation, Functional Correlation and Comparison with
Priya Narang1, Amar Agarwal2, Maheswari Srinivasan2
1Refractive & Glaucoma Services, Narang Eye Care & Laser Centre, Ahmedabad, India.
Ophthalmology Science
|October 17, 2022
Summary
The Advanced Vision Analyzer (AVA) demonstrates functional equivalence to the Humphrey Field Analyzer (HFA), showing strong correlation in visual field testing. This suggests AVA provides accurate visual field assessments comparable to established methods.
Area of Science:
- Ophthalmology
- Medical Technology
- Visual Field Testing
Background:
- Visual field testing is crucial for diagnosing and monitoring glaucoma and other optic nerve disorders.
- The Humphrey Field Analyzer (HFA) with its Swedish Interactive Threshold Algorithm (SITA) is a widely used standard for visual field assessment.
- Newer devices aim to provide comparable or improved accuracy and efficiency in visual field analysis.
Purpose of the Study:
- To evaluate the performance of the Advanced Vision Analyzer (AVA) by Elisar Vision Technology.
- To compare the pointwise threshold sensitivity and functional correlation of the Elisar Standard Algorithm (ESA) with the SITA algorithm of the HFA.
Main Methods:
- A prospective, cross-sectional study involving 160 eyes (85 control, 75 glaucoma patients) for functional assessment.
- Participants underwent both AVA (ESA) and HFA (SITA Standard 24-2) testing.
- Analyses included intraclass correlation coefficient (ICC), Bland-Altman plots, and linear regression to compare threshold measurements, test-retest variability (TRV), and blind spot accuracy.
Main Results:
- The AVA test duration was comparable to the HFA (7.08 vs. 6.26 minutes, P=0.228).
- Sectoral mean sensitivity showed statistically significant differences between AVA and HFA in both healthy and glaucoma participants (P < 0.001).
- Pointwise threshold values demonstrated moderate to strong correlations (r=0.68-0.89), and overall mean sensitivity showed good agreement (ICC=0.893).
Conclusions:
- The Advanced Vision Analyzer effectively captures visual field threshold values.
- Adequate functional correlation between AVA (ESA) and HFA (SITA Standard) suggests substantial equivalence.
- AVA shows potential for accurate visual field assessment, comparable to the HFA.
Keywords:
AVAAdvanced vision analyzerESA, Elisar Standard AlgorithmElisarElisar StandardHFAHMD, head-mounted deviceHumphrey Field AnalyzerICC, intraclass correlation coefficientLOA, limits of agreementMB, mean biasMD, mean deviationMS, mean sensitivityPR, patient responsePSD, pattern standard deviationPerimetrySITA StandardSITA, Swedish Interactive Threshold AlgorithmVirtual reality perimeterVisual fieldZESTZEST, Zippy Estimation of Sequential Threshold