Impact of the Insoluble Gas Concentration on Measured Stroke Volume at Rest and Submaximal Exercise Using the Innocor
Eli Reynolds1, Karrie Curry1, Gregory Barton2
1Division of Cardiology, Department of Internal Medicine, University of Texas Southwestern Medical School, Dallas, TX.
Introduction:
The Innocor® device uses an insoluble gas (SF 6 ) to estimate lung volume and the rate of disappearance of a soluble gas (nitrous oxide) to measure pulmonary blood flow (PBF), which approximates cardiac output assuming no shunt. We sought to identify error in the measurement of the insoluble gas in an effort to reduce variation in Innocor® measurement.
Methods:
We enrolled 28 participants from the Dallas Heart Study (mean age, 63 yr; 57% men; 43% White). Stroke volume was measured at rest and at submaximal (20 and 40 W) exercise using both echocardiography (Philips iE33) and the Innocor® device. We defined a priori peak and equilibrium SF 6 measurement errors as greater or less than 20% of the mean observed value. Three Innocor measurements were obtained at rest ( n = 27) for a total of 81 measurements. Of these, 22% had SF 6 measurements that fell outside of the a priori range.
Results:
Resting Innocor® stroke volume measures with peak SF 6 measured above a priori range (>0.12%) was associated with larger stroke volumes compared with stroke volume measures without peak SF 6 error (101.4 [26.8] vs 64.9 [8.7] mL; P = 0.006) and overestimated stroke volume when compared with stroke volume by echo (101.4 [26.8] vs 59.9 [16.3] mL; P = 0.017). A similar pattern was observed at submaximal exercise. In contrast, there was no consistent association between variation in equilibrium SF 6 concentrations and measured stroke volume.
Conclusions:
Variability in peak SF 6 concentration is common while using the Innocor® device and results in overestimated stroke volume. These findings have implications for research protocols using this device.
Related Concept Videos
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac...
Regulation of Stroke Volume
Preload refers to the degree of stretch on the heart before it contracts. It's analogous to the stretching of a rubber band; the more it's stretched, the more forcefully it snaps back. This concept is encapsulated in the Frank-Starling law of the...
Exercise and Cardiac Output
Sustained exercise increases the muscles' oxygen demand, which can be...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...