Critical Incident Reports Related to Ventilator Use: Analysis of the Japan Quality Council National Database
Naomi Akiyama1, Shihoko Kajiwara1, Takahiro Tamaki2
1From the School of Nursing, Gifu University of Health Science.
Objective:
This study aimed to assess the factors associated with medical device incidents.
Methods:
In this mixed-methods study, we used incident reporting data from the Japan Council for Quality Health Care. Of the 232 medical device-related reports that were downloaded, 34 (14.7%) were ventilator-associated incidents. Data related to patients, situations, and incidents were collected and coded.
Results:
The frequencies of ventilator-associated accidents were 20 (58.8%) during the daytime and 14 (41.2%) during the night/early morning. Ventilator-associated accidents occurred more frequently in the hospital room (n = 22 [64.7%]) than in the intensive care unit (n = 4 [11.8%]). Problems with ventilators occurred in only 4 cases (11.8%); in most cases, medical professionals experienced difficulty with the use or management of ventilators (n = 30 [88.2%]), and 50% of them were due to misuse/misapplication of ventilators (n = 17 [50.0%]). Ventilator-associated accidents were caused by an entanglement of complex factors-hardware, software, environment, liveware, and liveware-liveware interaction. Communication and alarm-related errors were reported to be related, as were intuitiveness or complicated specifications of the device.
Conclusions:
Our study revealed that ventilator-associated accidents were caused by an entanglement of complex factors and were related to inadequate communication among caregivers and families. Moreover, alarms were overlooked owing to inattentiveness. Mistakes were generally caused by a lack of experience, insufficient training, or outright negligence. To reduce the occurrence of ventilator-associated accidents, hospital administrators should develop protocols for employment of new devices. Medical devices should be developed from the perspective of human engineering, which could be one of the systems approaches.
More Related Videos
09:36Halogenated Agent Delivery in Porcine Model of Acute Respiratory Distress Syndrome via an Intensive Care Unit Type Device
Published on: September 24, 2020
08:34Investigation into Deep Breathing through Measurement of Ventilatory Parameters and Observation of Breathing Patterns
Published on: September 16, 2019
Related Concept Videos
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include:
Acute Respiratory Failure-I
Definition: It is defined by specific criteria based on blood gas measurements. Hypoxemia happens when the partial pressure of oxygen (PaO2) falls below 60 mmHg. At the same time,...
Acute Respiratory Failure-V
Ensure that patients are monitored continuously for their response to therapy, including changes in...
Acute Respiratory Failure-III
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Types of Reports II: Incident or Occurrence Report
Purposes:
In the healthcare industry, reports play a crucial role in documenting incidents within an agency. The primary objective of these reports is to ensure patient safety, uphold the...