Reproducibility assessment of radiomics features in various ultrasound scan settings and different scanner vendors
Yunus Soleymani1, Amir Reza Jahanshahi2, Amin Pourfarshid3
1Department of Neuroscience and Addiction Studies, School of Advanced Technologies in Medicine, Tehran University of Medical Sciences, Tehran, Iran.
Introduction:
Radiomics in Ultrasound (US) imaging has been investigated for the prediction and prognosis of cancers. However, inter-scanner and intra-scanner variations may affect the reproducibility of radiomics results. This study aims to evaluate the reproducibility of US textural radiomics features across various scan settings and scanner vendors.
Materials And Methods:
US images in quality control (QC) phantom were obtained by three scanners (Philips, Samsung, and Siemens) with different scan settings and parameters. Circular regions of interest (ROIs) inside isoechoic, hypoechoic, and hyperechoic objects were manually delineated. Forty textural radiomics features were extracted from each ROI, and then the robust features that could distinguish different echogenic objects were obtained by the Mann-Whitney U test. Reproducibility of the robust radiomics features was assessed by the intraclass correlation coefficient (ICC) and coefficient of variation (CV); ICC>0.90 and %CV<20 were considered reproducible.
Results:
According to the Mann-Whitney U test results, ten robust features could differentiate the hypoechoic, and 15 robust features could differentiate the hyperechoic objects from the isoechoic objects (P<0.001). The total ICC of the robust features for each echogenic object was >0.95 in different scanners and scan settings. Four and seven features were individually reproducible (%CV < 20, ICC>0.90) in hypoechoic and hyperechoic objects, respectively. Also, four features seem reproducible by changing the ROI location across the horizontal and vertical lines for both convex and linear array transducers.
Conclusions:
Most of the US textural radiomics features in this study were not reproducible. However, several features showed high reproducibility at different scan settings and scanners. These features may also be reproducible when ROI size and location change slightly.
More Related Videos
09:02Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population
Published on: January 31, 2025
16:01An Experimental Protocol for Assessing the Performance of New Ultrasound Probes Based on CMUT Technology in Application to Brain Imaging
Published on: September 24, 2017
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Imaging Studies II: Ultrasonography